Source: Sensors (Basel, Switzerland), written by Emanuela Cutuli; Lorena Saitta; Giulia Valenti; Gianluca Cicala; Francesca Guarino; Maide Bucolo (October 24, 2026). AI-generated summary by biochip.com, published . Not independently reviewed.
Key takeaways
- The University of Catania team tested its device with HL60 cells and PMMA particles at five concentrations, up to 10^7 particles per milliliter.
- Transmitted light declined as particle concentration rose, and laser illumination distinguished particle types more clearly than LED illumination.
- The results do not establish clinical use or long-term monitoring in an operating microfluidic system.
Researchers at the University of Catania developed a micro-optofluidic device that estimates microparticle concentration without labels, sample removal, or external imaging. The chip sends light through a microchannel and measures how much reaches a detector, using the change in transmitted light as an indicator of particle concentration. Tests with biological cells and synthetic plastic particles showed that the signal fell as concentration increased. The approach could help monitor samples inside microfluidic devices that are difficult to examine with a conventional microscope.
Measuring particles with light
Imagine shining a flashlight through fog: the denser the fog, the less light makes it straight through. The device uses a related effect, as suspended particles scatter light traveling through the channel, reducing the amount collected by an optical fiber.
The team tested the platform with HL60 cells and PMMA microparticles, a type of plastic particle, at five concentration levels. The range included a particle-free sample and concentrations up to 10^7 particles per milliliter. Because the method is label-free, it does not require fluorescent or other tags to make the particles detectable.
Laser and LED illumination
The researchers compared coherent laser light with light from a light-emitting diode, or LED. Both sources produced concentration-dependent changes in the transmitted signal, but laser illumination showed greater sensitivity and more clearly distinguished between the biological cells and the synthetic particles.
Computer simulations based on ray tracing modeled how light traveled through the device and interacted with particles. The simulations predicted particle-dependent differences in the light collected by the fiber, in line with the experimental response. A statistical analysis also found that particle type, concentration, illumination source, and interactions among those factors significantly affected the detected signal.
Why This Matters
Many microfluidic systems need to track cells or particles without repeatedly removing samples. Off-chip measurements can interrupt an experiment and introduce extra handling, while microscopy can be difficult when a channel is embedded or lacks a clear optical path. Integrating light delivery and detection into the device offers another way to measure a sample in place.
The reported work demonstrates concentration-dependent optical responses for two kinds of microparticles under controlled test conditions. It does not establish a clinical application or show how the device performs during long-term monitoring in an operating microfluidic system. Further work could build on the integrated optical design to develop self-contained sensing platforms.
Topics: Microfluidics
Start here: Microfluidics explained
Related articles
- Three-Dimensional Characterization of AC Electrothermal Flow in Microfluidic Devices
- Fabrication and Applications of Microfluidic Devices: A Review
- Chip in a lab: Microfluidics for next generation life science research
- Microfluidic Light-Scattering Imaging Coupled with Deep Learning for Label-Free Single-Cell Classification of Lymphoma Cells